The assessment of diagnostic medical images using 10-bit grayscale – fact or myth?
 
More details
Hide details
 
Publication date: 2018-04-05
 
 
Pol J Radiol, 2018; 83: 127-132
 
KEYWORDS
ABSTRACT
Purpose:
The aim of the study was to find tests for verification of 10-bit grayscale support for radiological workstation hardware/operating system and for DICOM viewer. Also, we wanted to perform the tests on available workstations.

Material and methods:
The following tests were selected to verify 10-bit grayscale support: workstation hardware and operating system tests using DirectX SDK 10BitScanout10.exe and NEC 10bitdemo.exe programs, DICOM viewer tests using TG18-MP and TG18-QC patterns. The tests were performed on 14 radiological workstations in a local department of radiology and on 4 radiological workstations during the technical exhibition at the ECR (European Congress of Radiology) in 2017.

Results:
Only 3 out of 14 local workstations passed the hardware and operating system 10-bit grayscale support test. Only one workstation passed the DICOM viewer 10-bit grayscale support test.

Conclusions:
Hardware/operating system and DICOM viewer 10-bit grayscale support should be tested on every radio­logical workstation even if compatibility is stated by the manufacturer. Existing medical diagnostic display quality requirements should be adjusted so that 10-bit support for DICOM viewers is only a desirable option.

 
REFERENCES (16)
1.
Minister Zdrowia. Obwieszczenie Ministra Zdrowia z dnia 26 kwietnia 2013 r. w sprawie ogłoszenia jednolitego tekstu rozporządzenia Ministra Zdrowia w sprawie warunków bezpiecznego stosowania promieniowania jonizującego dla wszystkich rodzajów ekspozycji medycznej. Dz.U. z 2013 poz. 1015. Available from: http://dziennikustaw.gov.pl/DU... [in Polish].
 
2.
Minister Zdrowia. Rozporządzenie zmieniające rozporządzenie w sprawie warunków bezpiecznego stosowania promieniowania jonizującego dla wszystkich rodzajów ekspozycji medycznej. Dz.U. z 2015 poz. 2040. Available from:.
 
4.
Minister Zdrowia. Obwieszczenie Ministra Zdrowia z dnia 3 kwietnia 2017 r. w sprawie ogłoszenia jednolitego tekstu rozporządzenia Ministra Zdrowia w sprawie warunków bezpiecznego stosowania promieniowania jonizującego dla wszystkich rodzajów ekspozycji medycznej. Dz.U. z 2017 poz. 884. Available from: http://dziennikustaw.gov.pl/DU... [in Polish].
 
5.
Sekcja Inżynierii Klinicznej PLTR. Wymagania dotyczące stanowiska do interpretacji (stanowiska opisowego) dla radiologii cyfrowej. 2010. Available from: http://los.net.pl/download.php... [in Polish].
 
6.
EUREF. European protocol for the quality control of the physical and technical aspects of mammography screening. In: European guidelines for quality assurance in breast cancer screening and diagnosis Fourth edition. 2006. Available from: http://www.euref.org/downloads....
 
7.
DIN Deutsches Institut für Normung. DIN 6868-157 Image quality assurance in diagnostic X-ray departments – Part 157: X-ray ordinance acceptance and constancy test of image display systems in their environment. Available from: http://www.din.de/en/getting-i....
 
8.
The Institute of Physics and Engineering in Medicine. Report 91 Recommended Standards For The Routine Performance Testing of Diagnostic X-Ray Imaging Systems. Available from: https://www.ipem.ac.uk/Scienti....
 
9.
American Association of Physicists in Medicine (AAPM) Task Group 18. Assessment of display performance for medical imaging systems. 2005. Available from: http://www.aapm.org/pubs/repor....
 
10.
Samei E, Badano A, Chakraborty D, Compton K et al.: Assessment of display performance for medical imaging systems: Executive summary of AAPM TG18 report. Med. Phys, 2005; 32(4): 1205–25.
 
11.
Japan Industries Association of Radiological Systems (JIRA). JESRA-X-0093 Quality Assurance (QA) Guideline for Medical Imaging Display Systems. Available from: http://www.jira-net.or.jp/e/in....
 
12.
nVidia. 30-Bit Color Technology for NVIDIA® Quadro®. 2009. Available from: https://www.nvidia.com/docs/IO....
 
13.
nVidia. 10 and 12-bit grayscale technology. 2013. Available from: ttps://www.nvidia.com/content/quadro_fx_product_literature/TB-04631-001_v05.pdf.
 
14.
Advanced Micro Devices. AMD’s 10-bit Video Output Technology. 2008. Available from: https://www.amd.com/Documents/....
 
15.
Tabár LK, Nińo AS, Schreiber E et al: Three digital mammography display configurations: Observer performance in a pilot ROC Study. In: Martí J, Oliver A, Freixenet J, Martí R (eds.). Digital mammography. IWDM 2010. Lecture Notes in Computer Science, vol. 6136. Springer, Berlin, Heidelberg.
 
16.
Kimpe T, Tuytschaever T. Increasing the number of gray shades in medical display systems – how much is enough? J Digit Imaging 2007; 20: 422-432.
 
Journals System - logo
Scroll to top